EVIDENT SCIENTIFIC INC: Device Recall
Recall #Z-2416-2024 · 07/22/2024
Class II: Risk
Recall Details
- Recall Number
- Z-2416-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- EVIDENT SCIENTIFIC INC
- Status
- Ongoing
- Date Initiated
- 07/22/2024
- Location
- Waltham, MA, United States
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- N/A
Reason for Recall
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Product Description
SlideView VS200 slide viewer,
Distribution Pattern
US