Canadian Hospital Specialties Ltd.: Device Recall

Recall #Z-2408-2024 · 06/11/2024

Class II: Risk

Recall Details

Recall Number
Z-2408-2024
Classification
Class II
Product Type
Device
Recalling Firm
Canadian Hospital Specialties Ltd.
Status
Ongoing
Date Initiated
06/11/2024
Location
Oakville, Canada
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6300 units

Reason for Recall

A customer found product with incomplete seals of P/N 20400161S, lot #185198 which can lead to non-sterile product and patient safety concerns. CHS has validated this complaint.

Product Description

Product Name: HIGH FLOW INSUFFLATION TUBING (with Filter) Model/Catalog Number: 20400161S Product Description: Laparoscopic Insufflator

Distribution Pattern

US Nationwide distribution in the state of MA.