Bard Peripheral Vascular Inc: Device Recall

Recall #Z-0887-2026 · 11/06/2025

Class II: Risk

Recall Details

Recall Number
Z-0887-2026
Classification
Class II
Product Type
Device
Recalling Firm
Bard Peripheral Vascular Inc
Status
Ongoing
Date Initiated
11/06/2025
Location
Tempe, AZ, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1240

Reason for Recall

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Product Description

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Distribution Pattern

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

Other Recalls by Bard Peripheral Vascular Inc

View all recalls by Bard Peripheral Vascular Inc →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.