Dexcom, Inc.: Device Recall
Recall #Z-0318-2026 · 08/28/2025
Recall Details
- Recall Number
- Z-0318-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Dexcom, Inc.
- Status
- Ongoing
- Date Initiated
- 08/28/2025
- Location
- San Diego, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 333
Reason for Recall
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).
Product Description
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
Distribution Pattern
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.