Healgen Scientific Llc: Device Recall

Recall #Z-3303-2024 · 08/22/2024

Class II: Risk

Recall Details

Recall Number
Z-3303-2024
Classification
Class II
Product Type
Device
Recalling Firm
Healgen Scientific Llc
Status
Ongoing
Date Initiated
08/22/2024
Location
Houston, TX, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,636,994 Tests

Reason for Recall

The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.

Product Description

Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.

Distribution Pattern

U.S. Nationwide distribution in the states of CA, GA and TX.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.