Microcare Medical: Device Recall
Recall #Z-3209-2024 · 08/29/2024
Recall Details
- Recall Number
- Z-3209-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Microcare Medical
- Status
- Ongoing
- Date Initiated
- 08/29/2024
- Location
- Commerce City, CO, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1649
Reason for Recall
Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk
Product Description
Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.