Beckman Coulter, Inc.: Device Recall
Recall #Z-3180-2024 · 07/10/2024
Recall Details
- Recall Number
- Z-3180-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 07/10/2024
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 170 units
Reason for Recall
Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
Product Description
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.