B Braun Medical Inc: Device Recall

Recall #Z-3177-2024 · 08/15/2024

Class II: Risk

Recall Details

Recall Number
Z-3177-2024
Classification
Class II
Product Type
Device
Recalling Firm
B Braun Medical Inc
Status
Ongoing
Date Initiated
08/15/2024
Location
Bethlehem, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
142336 units

Reason for Recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Product Description

STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095W

Distribution Pattern

US Nationwide distribution.