Btt Health Gmbh: Device Recall

Recall #Z-3162-2024 · 08/01/2024

Class II: Risk

Recall Details

Recall Number
Z-3162-2024
Classification
Class II
Product Type
Device
Recalling Firm
Btt Health Gmbh
Status
Ongoing
Date Initiated
08/01/2024
Location
Inning, Germany
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
62 units

Reason for Recall

Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).

Product Description

Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Distribution Pattern

US Nationwide distribution in the states of Florida.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.