Siemens Healthcare Diagnostics, Inc.: Device Recall

Recall #Z-3102-2024 · 07/26/2024

Class II: Risk

Recall Details

Recall Number
Z-3102-2024
Classification
Class II
Product Type
Device
Recalling Firm
Siemens Healthcare Diagnostics, Inc.
Status
Ongoing
Date Initiated
07/26/2024
Location
East Walpole, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated

Reason for Recall

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

Product Description

Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 11206239

Distribution Pattern

Worldwide - US Nationwide distribution