Siemens Healthcare Diagnostics, Inc.: Device Recall
Recall #Z-3101-2024 · 07/26/2024
Class II: Risk
Recall Details
- Recall Number
- Z-3101-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Status
- Ongoing
- Date Initiated
- 07/26/2024
- Location
- East Walpole, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
Reason for Recall
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
Product Description
Atellica IM CA 19-9 (250 Test)- In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN:10995489
Distribution Pattern
Worldwide - US Nationwide distribution