Wipro GE Healthcare Private Ltd.: Device Recall

Recall #Z-3081-2024 · 08/09/2024

Class II: Risk

Recall Details

Recall Number
Z-3081-2024
Classification
Class II
Product Type
Device
Recalling Firm
Wipro GE Healthcare Private Ltd.
Status
Ongoing
Date Initiated
08/09/2024
Location
Bengaluru, India
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
137 units

Reason for Recall

GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.

Product Description

GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.

Distribution Pattern

US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.