Wipro GE Healthcare Private Ltd.: Device Recall
Recall #Z-3081-2024 · 08/09/2024
Recall Details
- Recall Number
- Z-3081-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Wipro GE Healthcare Private Ltd.
- Status
- Ongoing
- Date Initiated
- 08/09/2024
- Location
- Bengaluru, India
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 137 units
Reason for Recall
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Product Description
GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; 4) M1112198-01047271; 5) M1112198-01048731; 6) M1112198-01048740; 7) M1112198-01052386; 8) M1112198-01054697; 9) M1112198-01068942; 10) M1112198-01084972; 11) M1112198-01090848; 12) M1112198-01093457; 13) M1112198-01117766; 14) M1112198-01139893; 15) M1112198-01156745; 16) M1112198-01167230; 17) M1112198-01177090; 18) M1112198-01183599; 19) M1112198-01185372; 20) M1112198-01193801; 21) M1112198-01200764; 22) M1112198-01245548; 23) M1112198-01253742; 24) M1112198-01274059. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Distribution Pattern
US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.