Siemens Healthcare Diagnostics, Inc.: Device Recall
Recall #Z-3065-2024 · 07/30/2024
Class II: Risk
Recall Details
- Recall Number
- Z-3065-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Status
- Ongoing
- Date Initiated
- 07/30/2024
- Location
- Newark, DE, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,371 units (4,394 US, 977 OUS)
Reason for Recall
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Product Description
Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Distribution Pattern
Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.