Mindray DS USA, Inc. dba Mindray North America: Device Recall

Recall #Z-2950-2024 · 07/22/2024

Class II: Risk

Recall Details

Recall Number
Z-2950-2024
Classification
Class II
Product Type
Device
Recalling Firm
Mindray DS USA, Inc. dba Mindray North America
Status
Ongoing
Date Initiated
07/22/2024
Location
Mahwah, NJ, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
96 units

Reason for Recall

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Product Description

Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00

Distribution Pattern

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.