Stryker Corporation: Device Recall
Recall #Z-2727-2024 · 07/24/2024
Recall Details
- Recall Number
- Z-2727-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Stryker Corporation
- Status
- Completed
- Date Initiated
- 07/24/2024
- Location
- Portage, MI, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2670 units
Reason for Recall
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Product Description
Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-030DC
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.