Beckman Coulter Mishima K.K.: Device Recall
Recall #Z-2691-2024 · 07/10/2024
Class II: Risk
Recall Details
- Recall Number
- Z-2691-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter Mishima K.K.
- Status
- Ongoing
- Date Initiated
- 07/10/2024
- Location
- Sunto-Gun, Japan
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 24 analyzers
Reason for Recall
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Product Description
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Distribution Pattern
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.