Beckman Coulter Mishima K.K.: Device Recall

Recall #Z-2691-2024 · 07/10/2024

Class II: Risk

Recall Details

Recall Number
Z-2691-2024
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter Mishima K.K.
Status
Ongoing
Date Initiated
07/10/2024
Location
Sunto-Gun, Japan
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24 analyzers

Reason for Recall

Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.

Product Description

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Distribution Pattern

Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.