Siemens Medical Solutions USA, Inc: Device Recall
Recall #Z-2664-2025 · 08/28/2025
Recall Details
- Recall Number
- Z-2664-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Status
- Ongoing
- Date Initiated
- 08/28/2025
- Location
- Malvern, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 152 units
Reason for Recall
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Product Description
MAGNETOM Verio. Model Number: 10276755.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Other Recalls by Siemens Medical Solutions USA, Inc
- Class II: Risk 12/29/2025
- Class II: Risk 12/29/2025
- Class II: Risk 12/19/2025
- Class II: Risk 12/19/2025
- Class II: Risk 12/19/2025