Covidien: Device Recall
Recall #Z-2637-2024 · 07/09/2024
Recall Details
- Recall Number
- Z-2637-2024
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Covidien
- Status
- Ongoing
- Date Initiated
- 07/09/2024
- Location
- Boulder, CO, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10,449 units
Reason for Recall
Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
Product Description
McGRATH" MAC EMS Video Laryngoscope, REF: 300-200-000 , Non-Sterile, Rx Only
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, Virgin Islands and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Bhutan, Bosnia, Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Cameroon, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican, Rep. Dutch Antilles, Egypt, El Salvador, Estonia, Faroe Islands, Fiji, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic of Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania Luxembourg Macao Malaysia Maldives Malta Martinique Mauritius Mexico Moldova, Republic of Mongolia, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russian Federation, Rwanda, Saint Martin, Saint Pierre And Miquelon, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Togo Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Viet Nam, Virgin Islands, U.S. Wallis And Futuna.
Other Recalls by Covidien
- Class II: Risk 06/23/2025
- Class I: Dangerous 05/21/2025
- Class I: Dangerous 02/26/2025
- Class I: Dangerous 07/09/2024
- Class II: Risk 07/09/2024