Philips North America Llc: Device Recall
Recall #Z-2595-2025 · 08/13/2025
Recall Details
- Recall Number
- Z-2595-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Philips North America Llc
- Status
- Ongoing
- Date Initiated
- 08/13/2025
- Location
- Cambridge, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 181 units
Reason for Recall
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Product Description
Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.