Ortho-Clinical Diagnostics, Inc.: Device Recall

Recall #Z-2559-2025 · 07/29/2025

Class II: Risk

Recall Details

Recall Number
Z-2559-2025
Classification
Class II
Product Type
Device
Recalling Firm
Ortho-Clinical Diagnostics, Inc.
Status
Ongoing
Date Initiated
07/29/2025
Location
Rochester, NY, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,908 units

Reason for Recall

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Product Description

Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.

Distribution Pattern

US Nationwide distribution.

Other Recalls by Ortho-Clinical Diagnostics, Inc.

View all recalls by Ortho-Clinical Diagnostics, Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.