Beckman Coulter, Inc.: Device Recall
Recall #Z-2545-2024 · 06/14/2024
Recall Details
- Recall Number
- Z-2545-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 06/14/2024
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 59 units
Reason for Recall
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
Product Description
Discrete photometric chemistry analyzer for clinical use.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.