Beckman Coulter, Inc.: Device Recall
Recall #Z-2544-2024 · 05/22/2024
Recall Details
- Recall Number
- Z-2544-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 05/22/2024
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 183 devices
Reason for Recall
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
Product Description
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Other Recalls by Beckman Coulter, Inc.
- Class II: Risk 12/23/2025
- Class II: Risk 11/26/2025
- Class II: Risk 11/26/2025
- Class II: Risk 11/10/2025
- Class II: Risk 11/07/2025