Max Mobility LLC: Device Recall
Recall #Z-2539-2025 · 08/12/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-2539-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Max Mobility LLC
- Status
- Ongoing
- Date Initiated
- 08/12/2025
- Location
- Lebanon, TN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15,834 units
Reason for Recall
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
Product Description
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
Distribution Pattern
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG
Other Recalls by Max Mobility LLC
- Class I: Dangerous 08/12/2025
- Class I: Dangerous 08/12/2025
- Class I: Dangerous 12/20/2024
- Class I: Dangerous 12/20/2024
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.