PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Device Recall

Recall #Z-2503-2025 · 08/01/2025

Class II: Risk

Recall Details

Recall Number
Z-2503-2025
Classification
Class II
Product Type
Device
Recalling Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Status
Ongoing
Date Initiated
08/01/2025
Location
Best, N/A, Netherlands
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,006 units

Reason for Recall

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

Product Description

Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.