American Contract Systems Inc: Device Recall

Recall #Z-2467-2025 · 07/14/2025

Class II: Risk

Recall Details

Recall Number
Z-2467-2025
Classification
Class II
Product Type
Device
Recalling Firm
American Contract Systems Inc
Status
Ongoing
Date Initiated
07/14/2025
Location
Tiffin, OH, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6617 units

Reason for Recall

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Product Description

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208439, Model/Item Number ANKP15AD; 3) LAPCHOLAPPY PACK, Model/Item Number FYLA38G; 4) CYSTO PACK, Model/Item Number HSCY13M; 5) PICC DRESSING CHANGE KIT, Model/Item Number HSDR10K; 6) MIDLINE CHANGE KIT, Model/Item Number HSMD20H; 7) SHOULDER ARTHOSCOPY PACK, Model/Item Number PDSA12D; 8) PDA LF SJW, Model/Item Number SWPD12K; 9) AAA PACK - 206038, Model/Item Number UDAA63AH; 10) BASIC CAB PACK - 206043, Model/Item Number UDBC35AF; 11) BASIC HEART PACK - 206044, Model/Item Number UDBH37AX; 12) FREE FLAP DRAPE PACK (PS 907864), Model/Item Number UIFF72AD; 13) PROCUREMENT PACK, Model/Item Number UIPC89D; 14) THYROID PACK (PS 096552), Model/Item Number UITR45W; 15) TRAUMA PACK 175634, Model/Item Number UMTR17B; 16) TRAUMA PACK 175634, Model/Item Number UMTR17C.

Distribution Pattern

US distribution to states of: IL and LA