BioFire Diagnostics, LLC: Device Recall

Recall #Z-2456-2024 · 06/19/2024

Class II: Risk

Recall Details

Recall Number
Z-2456-2024
Classification
Class II
Product Type
Device
Recalling Firm
BioFire Diagnostics, LLC
Status
Ongoing
Date Initiated
06/19/2024
Location
Salt Lake City, UT, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
99,822

Reason for Recall

If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.

Product Description

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.