VANTIVE US HEALTHCARE LLC: Device Recall
Recall #Z-2439-2025 · 07/28/2025
Class II: Risk
Recall Details
- Recall Number
- Z-2439-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- VANTIVE US HEALTHCARE LLC
- Status
- Ongoing
- Date Initiated
- 07/28/2025
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7432 units
Reason for Recall
Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted
Product Description
TherMax Blood Warmer Unit-US, PrisMax Accessory, Product Codes: 955630, 955515, 955702, 955631
Distribution Pattern
Worldwide
Other Recalls by VANTIVE US HEALTHCARE LLC
- Class II: Risk 01/06/2026
- Class II: Risk 01/06/2026
- Class II: Risk 01/06/2026
- Class II: Risk 01/06/2026
- Class II: Risk 01/06/2026
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.