Boston Scientific Corporation: Device Recall
Recall #Z-2429-2025 · 07/29/2025
Recall Details
- Recall Number
- Z-2429-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Boston Scientific Corporation
- Status
- Ongoing
- Date Initiated
- 07/29/2025
- Location
- Maple Grove, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 340185 units in total
Reason for Recall
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
Product Description
WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.
Distribution Pattern
Worldwide.