Radiometer Medical ApS: Device Recall

Recall #Z-2426-2024 · 05/31/2024

Class II: Risk

Recall Details

Recall Number
Z-2426-2024
Classification
Class II
Product Type
Device
Recalling Firm
Radiometer Medical ApS
Status
Ongoing
Date Initiated
05/31/2024
Location
Bronshoj, Denmark
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27,127 analyzers

Reason for Recall

Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.

Product Description

ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) " In vitro testing of samples of expired air for the parameters pO2 and pCO2 " In vitro testing of pleura samples for the pH parameter.

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.