Dentsply IH, Inc.: Device Recall
Recall #Z-2422-2025 · 07/24/2025
Recall Details
- Recall Number
- Z-2422-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Dentsply IH, Inc.
- Status
- Ongoing
- Date Initiated
- 07/24/2025
- Location
- Waltham, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 780 units
Reason for Recall
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
Product Description
Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
Distribution Pattern
Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
Other Recalls by Dentsply IH, Inc.
- Class II: Risk 07/24/2025