Dentsply IH, Inc.: Device Recall

Recall #Z-2421-2025 · 07/24/2025

Class II: Risk

Recall Details

Recall Number
Z-2421-2025
Classification
Class II
Product Type
Device
Recalling Firm
Dentsply IH, Inc.
Status
Ongoing
Date Initiated
07/24/2025
Location
Waltham, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
102 units

Reason for Recall

The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.

Product Description

Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component

Distribution Pattern

Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.