Beckman Coulter, Inc.: Device Recall
Recall #Z-2414-2024 · 06/04/2024
Class II: Risk
Recall Details
- Recall Number
- Z-2414-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 06/04/2024
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 143 devices
Reason for Recall
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
Product Description
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.