Ortho-Clinical Diagnostics, Inc.: Device Recall

Recall #Z-2335-2025 · 07/09/2025

Class II: Risk

Recall Details

Recall Number
Z-2335-2025
Classification
Class II
Product Type
Device
Recalling Firm
Ortho-Clinical Diagnostics, Inc.
Status
Ongoing
Date Initiated
07/09/2025
Location
Rochester, NY, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
174,215 units

Reason for Recall

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.

Product Description

VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

Other Recalls by Ortho-Clinical Diagnostics, Inc.

View all recalls by Ortho-Clinical Diagnostics, Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.