Barco N.V.: Device Recall

Recall #Z-2332-2024 · 04/24/2024

Class II: Risk

Recall Details

Recall Number
Z-2332-2024
Classification
Class II
Product Type
Device
Recalling Firm
Barco N.V.
Status
Completed
Date Initiated
04/24/2024
Location
Kortrijk, N/A, Belgium
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
88 units

Reason for Recall

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Product Description

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Distribution Pattern

US: FL