Barco N.V.: Device Recall
Recall #Z-2332-2024 · 04/24/2024
Class II: Risk
Recall Details
- Recall Number
- Z-2332-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Barco N.V.
- Status
- Completed
- Date Initiated
- 04/24/2024
- Location
- Kortrijk, N/A, Belgium
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 88 units
Reason for Recall
We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
Product Description
MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display
Distribution Pattern
US: FL
Other Recalls by Barco N.V.
- Class II: Risk 05/09/2025
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.