Boston Scientific Corporation: Device Recall
Recall #Z-2325-2025 · 07/24/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-2325-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Boston Scientific Corporation
- Status
- Ongoing
- Date Initiated
- 07/24/2025
- Location
- Saint Paul, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 247,702 US; 344,670 OUS
Reason for Recall
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Product Description
ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
Distribution Pattern
Worldwide Distribution. US nationwide. International distribution worldwide.