Boston Scientific Corporation: Device Recall

Recall #Z-2325-2025 · 07/24/2025

Class I: Dangerous

Recall Details

Recall Number
Z-2325-2025
Classification
Class I
Product Type
Device
Recalling Firm
Boston Scientific Corporation
Status
Ongoing
Date Initiated
07/24/2025
Location
Saint Paul, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
247,702 US; 344,670 OUS

Reason for Recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Product Description

ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187

Distribution Pattern

Worldwide Distribution. US nationwide. International distribution worldwide.