American Contract Systems Inc: Device Recall
Recall #Z-2284-2024 · 05/15/2024
Class II: Risk
Recall Details
- Recall Number
- Z-2284-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- American Contract Systems Inc
- Status
- Ongoing
- Date Initiated
- 05/15/2024
- Location
- Tiffin, OH, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 552 units
Reason for Recall
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
Product Description
BPKN15C KNEE BASIN PACK BHS- Procedure tray Catalog Number: BPKN15C
Distribution Pattern
US Nationwide distribution.