Siemens Healthcare Diagnostics Inc: Device Recall

Recall #Z-2272-2025 · 07/07/2025

Class II: Risk

Recall Details

Recall Number
Z-2272-2025
Classification
Class II
Product Type
Device
Recalling Firm
Siemens Healthcare Diagnostics Inc
Status
Ongoing
Date Initiated
07/07/2025
Location
Norwood, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37,554 units

Reason for Recall

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.

Product Description

epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.

Other Recalls by Siemens Healthcare Diagnostics Inc

View all recalls by Siemens Healthcare Diagnostics Inc →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.