Siemens Healthcare Diagnostics Inc: Device Recall
Recall #Z-2272-2025 · 07/07/2025
Recall Details
- Recall Number
- Z-2272-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Siemens Healthcare Diagnostics Inc
- Status
- Ongoing
- Date Initiated
- 07/07/2025
- Location
- Norwood, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 37,554 units
Reason for Recall
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
Product Description
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.
Other Recalls by Siemens Healthcare Diagnostics Inc
- Class II: Risk 10/06/2025
- Class II: Risk 10/06/2025
- Class II: Risk 10/06/2025
- Class II: Risk 09/22/2025
- Class II: Risk 08/11/2025