Beckman Coulter, Inc.: Device Recall
Recall #Z-2271-2025 · 06/25/2025
Class II: Risk
Recall Details
- Recall Number
- Z-2271-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 06/25/2025
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 177 systems
Reason for Recall
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Product Description
DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460
Distribution Pattern
Worldwide distribution - US Nationwide.