Getinge Usa Sales Inc: Device Recall

Recall #Z-2271-2024 · 05/29/2024

Class II: Risk

Recall Details

Recall Number
Z-2271-2024
Classification
Class II
Product Type
Device
Recalling Firm
Getinge Usa Sales Inc
Status
Ongoing
Date Initiated
05/29/2024
Location
Wayne, NJ, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
73 units

Reason for Recall

The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.

Product Description

ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B4

Distribution Pattern

Domestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech Republic, Denmark, France, Greece, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Mexico, New Zealand, Norway, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.

Other Recalls by Getinge Usa Sales Inc

View all recalls by Getinge Usa Sales Inc →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.