Beckman Coulter Inc.: Device Recall

Recall #Z-2264-2025 · 06/26/2025

Class II: Risk

Recall Details

Recall Number
Z-2264-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter Inc.
Status
Ongoing
Date Initiated
06/26/2025
Location
Brea, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
56,696 units

Reason for Recall

Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

Product Description

UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

Distribution Pattern

US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nigeria, North Macedonia, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Province of China, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen.

Other Recalls by Beckman Coulter Inc.

View all recalls by Beckman Coulter Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.