CareFusion 303, Inc.: Device Recall
Recall #Z-2255-2024 · 05/13/2024
Recall Details
- Recall Number
- Z-2255-2024
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- CareFusion 303, Inc.
- Status
- Ongoing
- Date Initiated
- 05/13/2024
- Location
- San Diego, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 677
Reason for Recall
Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.
Product Description
BD Pyxis, Name/REF: MEDBANK MINI 1FH-1FM/ 169-137, MEDBANK MINI 1FH-2HH MEDPASS/ 169-151, MEDBANK MINI 1HH-1FH-1HM/ 169-101, MEDBANK MINI 1HH-1HM-1FM/ 169-102, MEDBANK MINI 1HH-3HM/ 169-103, MEDBANK MINI 2FH/ 169-104, MEDBANK MINI 2FM/ 169-105, MEDBANK MINI 2HH-1FH/ 169-106, MEDBANK MINI 2HH-1FM/ 169-107, MEDBANK MINI 2HH-1FM MEDPASS/ 169-152, MEDBANK MINI 2HH-2HM/ 169-108, MEDBANK MINI 2HM-1FM/ 169-109, MEDBANK MINI 3HH-1FG/ 169-110, MEDBANK MINI 3HH-1HM/ 169-111, MEDBANK MINI 4HH/ 169-112, MEDBANK MINI 4HH MEDPASS/ 169-150, MEDBANK MINI 4HM/ 169-113, MEDBANK MINI CR-2HH-1FH-P/ 169-114, MEDBANK MINI CR-4HH-P/ 169-115, MB MINI MAIN CR/ 169-163, MB MINI MAIN W/CUBIE DRWS/ 169-159, MB MINI MAIN W/FH FALSE DRW/ 169-162, MB MINI MAIN W/HH FALSE DRW/ 169-161, MB MINI MAIN W/MATRIX DRWS / 169-160, MB MINI MAIN W/MP/ 169-164 Used with the following user guides: BD Pyxis CUBIE Replenishment Station User Guide (DIR 10000414989), BD Pyxis MedBank Cabinet User Guide (DIR 10000414987), BD Pyxis MedBank Installation and Upgrade Guide (DIR 10000414990)
Distribution Pattern
US Nationwide distribution including in the states of NY, CA, MO, AZ, CO, KS, IL, FL, OK, OH, TN, GA, NM, NV, NE, TX, NC, KY, WI, SC, IA, MT, SD, VA, NJ, PA, AK, MS, MI, WA, UT, ND, MA, MN, OR, IN, WV, AL, ID, LA, CT, WY, MD, NH, AR, HI, DC.
Other Recalls by CareFusion 303, Inc.
- Class II: Risk 12/03/2025
- Class II: Risk 11/20/2025
- Class II: Risk 11/06/2025
- Class I: Dangerous 10/17/2025
- Class II: Risk 09/30/2025