Ortho-Clinical Diagnostics, Inc.: Device Recall

Recall #Z-2245-2024 · 05/14/2024

Class III: Low Risk

Recall Details

Recall Number
Z-2245-2024
Classification
Class III
Product Type
Device
Recalling Firm
Ortho-Clinical Diagnostics, Inc.
Status
Ongoing
Date Initiated
05/14/2024
Location
Rochester, NY, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1718

Reason for Recall

Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.

Product Description

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344

Distribution Pattern

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Other Recalls by Ortho-Clinical Diagnostics, Inc.

View all recalls by Ortho-Clinical Diagnostics, Inc. →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.