Siemens Healthcare Diagnostics Inc: Device Recall

Recall #Z-2214-2024 · 05/17/2024

Class II: Risk

Recall Details

Recall Number
Z-2214-2024
Classification
Class II
Product Type
Device
Recalling Firm
Siemens Healthcare Diagnostics Inc
Status
Ongoing
Date Initiated
05/17/2024
Location
Norwood, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units US

Reason for Recall

Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.

Product Description

Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom.