CareFusion 303, Inc.: Device Recall
Recall #Z-2179-2025 · 07/08/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-2179-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- CareFusion 303, Inc.
- Status
- Ongoing
- Date Initiated
- 07/08/2025
- Location
- San Diego, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 33,162,680 units total
Reason for Recall
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Product Description
BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
Distribution Pattern
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
Other Recalls by CareFusion 303, Inc.
- Class II: Risk 12/03/2025
- Class II: Risk 11/20/2025
- Class II: Risk 11/06/2025
- Class I: Dangerous 10/17/2025
- Class II: Risk 09/30/2025