Baxter Healthcare Corporation: Device Recall

Recall #Z-2173-2025 · 07/14/2025

Class I: Dangerous

Recall Details

Recall Number
Z-2173-2025
Classification
Class I
Product Type
Device
Recalling Firm
Baxter Healthcare Corporation
Status
Ongoing
Date Initiated
07/14/2025
Location
Deerfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36705 units

Reason for Recall

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.

Product Description

Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.