Recall Number
Z-2157-2025
Classification
Class II
Product Type
Device
Recalling Firm
Bridge to Life Ltd
Status
Ongoing
Date Initiated
06/12/2025
Location
Northbrook, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4528 units

Reason for Recall

The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.

Product Description

Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250

Distribution Pattern

United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada