Bridge to Life Ltd: Device Recall
Recall #Z-2157-2025 · 06/12/2025
Recall Details
- Recall Number
- Z-2157-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Bridge to Life Ltd
- Status
- Ongoing
- Date Initiated
- 06/12/2025
- Location
- Northbrook, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4528 units
Reason for Recall
The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
Product Description
Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250
Distribution Pattern
United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada