Olympus Corporation of the Americas: Device Recall

Recall #Z-2140-2025 · 06/18/2025

Class II: Risk

Recall Details

Recall Number
Z-2140-2025
Classification
Class II
Product Type
Device
Recalling Firm
Olympus Corporation of the Americas
Status
Ongoing
Date Initiated
06/18/2025
Location
Center Valley, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
41 units

Reason for Recall

Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

Product Description

Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

Distribution Pattern

International distribution to the country of Japan.