Blue Belt Technologies, Inc: Device Recall

Recall #Z-2135-2024 · 05/07/2024

Class II: Risk

Recall Details

Recall Number
Z-2135-2024
Classification
Class II
Product Type
Device
Recalling Firm
Blue Belt Technologies, Inc
Status
Ongoing
Date Initiated
05/07/2024
Location
Pittsburgh, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
916 units

Reason for Recall

CORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to flicker on screen of the CORI Surgical System during use, and may cause surgical delay

Product Description

CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures Product Numbers: (1) ROB10000 CORI Starter Kit (2) ROB10024 Real Intelligence CORI (3) ROB10197 CORI STARTER KIT - UNIVERSAL Note: the three product numbers cover the three different configurations in which the CORI Console is sold

Distribution Pattern

Nationwide Foreign: Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK