GE Medical Systems, LLC: Device Recall

Recall #Z-2132-2025 · 05/23/2025

Class II: Risk

Recall Details

Recall Number
Z-2132-2025
Classification
Class II
Product Type
Device
Recalling Firm
GE Medical Systems, LLC
Status
Ongoing
Date Initiated
05/23/2025
Location
Waukesha, WI, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Reason for Recall

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Product Description

Discovery MR750w 3.0T

Distribution Pattern

Worldwide distribution - US Nationwide and the country of China.