GE Medical Systems, LLC: Device Recall
Recall #Z-2131-2025 · 05/23/2025
Class II: Risk
Recall Details
- Recall Number
- Z-2131-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- GE Medical Systems, LLC
- Status
- Ongoing
- Date Initiated
- 05/23/2025
- Location
- Waukesha, WI, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Reason for Recall
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Product Description
SIGNA Architect AIR
Distribution Pattern
Worldwide distribution - US Nationwide and the country of China.